Number of Channels
Four (4).
Memory technology
EEProm; memory backup power not required, over 200 year data
retention.
Inputs Accepted
All PACE sensors and probes, 0-5vdc, 2-wire
resistance (30k ohm midpoint), 3-wire potentiometer (>1k ohm),
switch closure. Inputs auto-configure for sensor or input range
selected in software. Sensors, probes and
input signal types can be mixed in any
combination, except PM-1
Pulse Module which is available on
channel 4 only.
Additional input ranges of +/-5vdc, 0-10vdc, 0-20vdc, 0-30vdc, 0-60vdc and 4-20ma are available using the Input Scaling Module.
Additional input ranges of ±11mv, ±22mv, ±107mv, 0-20mv, 0-50mv, and 0-500mv are available using a Thermocouple / Millvolt Module.
Starting Modes (3)
Start immediate, Start at pre-set time and date, Start on Ch1 Trigger*.
Recording Modes (3)
Record continuously, Record until memory is full, Record until Ch1 Stop
Trigger*.
*If Channel 1 is set for a temperature probe or a switch closure (using Pocket Logger Software), the terminal block can be plugged in to start and/or unplugged to end a logging session (a switch closure can do the same). Ch1 may be set to 'OFF' to maximize memory capacity of 'ON' channels.
NOTE: The Start/Stop Trigger function is a 'one time only' start and/or a 'one time only' stop for a given logging session.
Sampling Rates (20)
2, 4, 6, 10, 20, 30 seconds; 1, 2, 5, 10, 15, 20, 30 minutes; 1,
2, 3, 4, 6, 8, 12 hours. Timing accuracy: ± 2
min. / month for 10 sec. - 12 hr. rates; Timing
accuracy for 2, 4, & 6 sec. rates is ±30% if memory is in
wrap-around mode, otherwise accuracy is ± 2 min. / month.
Fast Sampling Rates(8)
200/sec, 100/sec, 50/sec, 20/sec, 10/sec, 5/sec, 2/sec, 1/sec. Note:
While recording at fast sampling rates, communication with a PC
is not available until logging has stopped. Timing
accuracy: ± 0.5%.
Sampling Modes (3)
Single Point, Average (accumulate readings every 2 seconds;
compute and store average value at preset sample time), Pulse
Counting (channel 4 only, requires PM-1 Pulse Module).
Text blocks (5)
The XR440 Data Logger accepts custom labels for each channel and the
logging session. These labels display on the status screen, real
time screen, all created graphs and ASCII files.
Real Time Display Mode
XR440 input channel values (in engineering units) are
updated every 2 seconds (while connected to a PC running Pocket
Logger Software). Real Time Display Mode does not affect any
ongoing logging session.
Real Time Clock
Counts seconds, minutes, hours, day of week, month, date of
month, and year with leap year compensation.
Clock accuracy: Better than +/- 2 min. per month
at 25°C.
Battery
Replaceable 9 volt lithium battery. Battery life: Estimated 2 - 3
years. A standard 9 volt alkaline battery can also be used (est.
life: 15 months). Actual battery voltage displays on PC (when
logger is connected to a PC running Pocket Logger Software).
Enclosure
Impact-resistant ABS plastic case.
Size
4.70" by 2.40" by .95"
120 mm x 61 mm x 24 mm.
Weight
6 ounces; 156 grams (includes battery).
Operating ambient temperature range
-40°C to 60°C. (-40°F to 140°F).
Operating ambient humidity range
0 to 90% relative humidity (non-condensing).
Communication
Serial RS232C. Baud Rates: 1200, 2400, 4800, 9600 and 19.2k baud.
Data and status transfers to PC (and real time display mode) do
not affect any ongoing logging session.
Interface Cable
Terminations: 9 pin D-Sub female (connects to PC); 3.5mm stereo
phone plug (connects to Pocket Logger). Length: 6 feet; 1.8
meters. (Part No. IC209)
Recommended sensor cable
Two, three, or four conductor, 22 - 24 gage, stranded copper.
Maximum resistive probe cable length
Using 24 gage shielded cable: 500 feet (150 meters). Greater
cable lengths are possible using heavier gage cable.
Input connection
Removable screw-type terminal block.
Input impedance
Greater than 5 megohms (while sampling).
Input protection
Over/under voltage protection on all inputs: 40 amps peak 8/20
microseconds.
Approvals
CE approved to EN50081-1 (RF emissions) and
EN50082-1 (ESD and RF immunity) using any combination of CE
approved PACE sensors or probes with standard lead lengths. All
PACE Sensors and Probes are CE approved except the following:
PACE SC series AC Current Sensors, PACE SV series
AC Voltage
Sensors.
Calibration
Not required; Accuracy specifications are maintained without user
adjustment. Contains no adjustable components. An NIST traceable Validation
Certificate is included with each XR440 Data Logger. An
accuracy validation procedure is included in the Pocket Logger Software's online help.
Pulse Counting (using PM-1
Pulse Module):
Maximum Rate: 5 pulses per second
Minimum Closure: 30 milliseconds
Maximum Pulses (see below)
Resolution dependent specifications
The XR440 can be set to 12, 10 or 8 bit resolution. Selecting a lower resolution increases the number of readings which can be stored in the XR440's memory.
| 12 bit resolution | |
| Memory capacity | XR440: 21,504 readings XR440-M: 86,016 readings |
| Voltage input accuracy | ±0.25% F.S. (0-5 vdc) |
| Temperature accuracy (logger + PACE probe) | ±0.15°C from 0 to 40°C (±0.27°F from 32 to 104°F) |
| ±0.3°C from -25 to 85°C (±0.55°F from -13 to 185°F) | |
| Resolution | 12 bits; 0.025% of F.S. (0-5vdc) |
| 0.025°C from 0 to 40°C | |
| 0.063°C or better -25 to 75°C | |
| 0.125°C or better -45 to 100°C | |
| Max pulses / sample interval | 4095 (with PM-1 Pulse Module) |
| 10 bit resolution | |
| Memory capacity | XR440: 25,800 readings XR440-M: 103,200 readings |
| Voltage input accuracy | ±0.32% F.S. (0-5 vdc) |
| Temperature accuracy (logger + PACE probe) | ±0.22°C from 0 to 40°C (±0.4°F from 32 to 104°F) |
| ±0.5°C from -25 to 85°C (±0.9°F from -13 to 185°F) | |
| Resolution | 10 bits; 0.1% of F.S. (0-5 vdc) |
| 0.1°C from 0 to 40°C | |
| 0.25°C or better -25 to 75°C | |
| 0.5°C or better -45 to 100°C | |
| Max pulses / sample interval | 1023 (with PM-1 Pulse Module) |
| 8 bit resolution | |
| Memory capacity | XR440: 32,256 readings XR440-M: 129,024 readings |
| Voltage input accuracy | ±0.65% F.S. (0-5 vdc) |
| Temperature accuracy (logger + PACE probe) | ±0.65°C from 0 to 40°C (±1.2°F from 32 to 104°F) |
| ±1.0°C from -25 to 75°C (±1.8°F from -13 to 167°F) | |
| Resolution | 8 bits; 0.4% of F.S. (0-5 vdc) |
| 0.4°C from 0 to 40°C | |
| 1°C or better -25 to 75°C | |
| 2°C or better -45 to 100°C | |
| Max pulses / sample interval | 255 (with PM-1 Pulse Module) |
What does 8, 10 and 12 bit mean?